Description
|
APPLICATION DESCRIPTION
|
||||||||
Specifications
OUTPUT PULSE
Charging voltage: Adjustable from 0 to ≈1900 V
Rise/fall time: <2 ns
Duration time: 35 ns
Pulse repetition rate: <1.0 pulse per second to 120 pulse per second
Polarity: ± selectable
Output load: 50 ± j 0 Ω
Dimensions: 12.25″ wide x 8.7″ high x 13″ deep (31.1 cm x 21.1 cm x 33.0 cm)
Weight: 27 pounds (12.24 kg)
Shipping weight: 30 pounds (13.60 kg)
Useful Accessories
ACCESSORIES RECOMMENDED FOR CS115 TESTING
Solar Type 9233-50-TS-50-N Line Impedance Stabilization Network
Solar Type 9125-1 Calibration Fixture
Used to calibrate probes with a window size from 1.25″ to 1.50″ and a frequency range of 20 Hz to 500 MHz
Solar Type 9142-1N Current Injection Probe
with a frequency range of 2 MHz to 500 MHz, 200 W
Solar Type 9123-1N Current Monitor Probe
with a frequency range of 10 kHz to 500 MHz, 1.25″ window
Solar Type 9410-1 High Voltage 40 dB Attenuator
The Type 9410-1 high voltage 50 ohm attenuator has an insertion loss of 40 dB ± 1.5 dB from 100 Hz to 100 MHz and 40 dB ± 3 dB from 100 MHz to 1 GHz and an average wattage dissipation capability of 2 watts. The attenuator will reduce the pulse voltage by a factor of 100 to 1 (40 dB). 250 volts will be developed when 5 amperes is flowing through the 50 ohm input of the attenuator. Therefore, once the attenuation factor is applied, the voltage present at the output of the attenuator will be 2.5 volts. (250 volts – 40 dB = 2.5 volts). See your oscilloscope manual for more details on 50 ohm input.
Solar Type 9841-1 1000 V Termination
50 Ω coaxial 1 W average power. Typical input VSWR in a 50 Ω system under 1.5 from DC to 1 GHz